Patient Monitor System
MM-C003 Patient Monitor
Product Description
Features :
- Standard parameter: ECG,RESP,NIBP, SPO2, TEMP, PR
- 12.1 large-screen color TFT display
- Built in high capacity battery, up to 4 hours working tine
- 7- lead ECG waveforms display , big fonts observational interface
- Reviewable ECG waveforms, maximum 240 hours reviewing graphic trends
- Digital SPO2 technique, precise measurement for weal perfusion and finger shaking
- Arrhythmia and S-T segment analysis and pacemaker detection
- Protection against interference from defibrillation and electrosurgical unit
- Applicable to ICU, CCU, operation, adult, pediatric and neonatal patients
Specification :
- ECG
- Full lead: I, II, III, AVL, AVR, AVF and V
- Gain 5mm/mV, 10mm/mV, 15mm/mV, 20mm/mV and 25mm/mV
- Scanning speed: Changeable with three grades of 12.5mm/s, 25mm/s and 50mm/s
- Band width :0.05Hz-100Hz(3dB)
- Heart rate:
- Scope: 15 times/minute to 300 times/minutes
- Accuracy: ±1 time/minute
- Variation of heart beat LORENZE drawing: Analysis on heart beat data for the current 5 minutes with histogram during RR period.
- Arrhythmia :Twenty kinds of common abnormal ECG can be analyzed, with ECG data of 1000 times of arrhythmia kept and with 10 seconds kept for each time.
- Accuracy: Error rate < 20%
- Undetected rate < 20%
- Analysis function should be started when using the above two kinds of function.
- ST section
- Scope: ±0.8mV
- Precision : ±0.05mV
SpO2
- Detection scope: 0%-100%
- Accuracy : ±1% 90%-100% ±2% 70%-89%
NIBP
- Theory Electric oscillator
- Cuff gassing time less than 15 seconds
- First-time gassing pressure: dynamic state 24kPa 180 mmHg
- Manual selection: 9.3kPa 70 mmHg13.3kPa 100 mmHg16.0kPa 120 mmHg
- 18.6kPa 140 mmHg20.0kPa 150 mmHg21.3kPa 160 mmHg
- Automatic detection period: 1 minute to 240 minutes
- Stop size1 min1 min -10min, 5min(10 min -30min), 10min(30 min -90min) and 30min90 min -240min
- Detection scope:
- Systolic blood pressure: 4kPa -34kPa(30mmHg -255mmHg)
- Diastolic pressure: 2kPa -29.3kPa(15mmHg -220mmHg)
- Mean pressure : 2.7kPa31.3kPa 20mmHg235mmHg
- Detection accuracy
- Static pressure: ±3mmHg
- Pulse rate: ±2%
Resp
- Chest impedance method or concentration of carbon dioxide at end of respiration
- Scope of chest impedance method 0 time/minute – 100 minutes/minute
- Precision of chest impedance method 1 time/minute
- Detection scope of carbon oxygen at end of respiration 3mmHg-75mmHg 0.4%vol-9.9%vol
- Detection scope of respiration rate: 2 times/minute ->60 times/minute including baby
- Detection precision of respiration rate: ±1 time/minute
- Temp
- Single/double path
- Type : Vascular/surface temperature
- Scope: 20C50C
- Accuracy : ±0.2 C
Invasive blood pressure (selected mating)
- Single/double path
- Type: direct and continuous detection
- Scope: -6.67kPa(-50mmHg)—46.67kPa(350mmHg)
- Precision: below 13.33kPa(100mmHg) ±0.27kPa(2mmHg) 13.33kPa(100mmHg) above ±2%
- Frequency responseDC-12Hz
- Input resistance: not less than 500kΩ
- Sensitivity of sensor: 5μV/V/mmHg
- Motivation voltage5V (DC)
- Accuracy of sensor: ±2
- Isolation voltage: 4 kV
Invasive blood pressure (selected mating)
- Single/double path
- Type: direct and continuous detection
- Scope: -6.67kPa(-50mmHg)—46.67kPa(350mmHg)
- Precision: below 13.33kPa(100mmHg) ±0.27kPa(2mmHg) 13.33kPa(100mmHg) above ±2%
- Frequency response:DC-12Hz
- Input resistance: not less than 500kΩ
- Sensitivity of sensor: 5μV/V/mmHg
- Motivation voltage 5V (DC)
- Accuracy of sensor: ±2
- Isolation voltage: 4 kV
|