Patient Monitor System
MM-C004 Modular Type Patient Monitor
Product Description
Features :
- 6 Standard parameters: ECG, RESP, NIBP, SPO2, 2-TEMP, PR/HR, ETCO2, 2IBP and Recorder, AG, BIS, CO, OB/GYN (TOCO,FHR, FM) are optional
- Patient range: Neonate, Pediatric, and adult patients
- Touch Screen 15"high resolution 1024 X 768 color TFT LCD display
- Real time S-T segment analysis (15 types), drug calculation (15 types), pace-maker detection.
- Indication of Incorrect manipulation and malfunction analysis.
- Patient info input management function
- Multi-lead ECG waveforms display in phase.
- Large volume of tabular and graphic trends information easy to recall.
- Capture dynamic waveforms.
- Efficient resistance to interference of defibrillator and HF knife.
- UP to 6 hours working capacity of built-in rechargeable 2 battery.
Performance Specifications :
Display :
- 15"color TFT Touch Screen
- Rolling and refreshing waveform display
- Resolution: 1024x768
- Standard display
- Multi channels display
Trace :
- 6 waveforms
- Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
Indicator :
- Power/battery indicator light
- QRS beep and alarm sound
Battery :
- Rechargeable lead acid cell, 12V/4AH
- Max.24 hours for charging, 3 hours for continuous working
Trend time :
Alarm :
- User-adjustable High , Medium and Low limits 3-level Audible and visual alarm
Recorder :
- Built-in , thermal array
- Plethysmogram waveform: 2 channels
- Record mode: manual, on alarm, time-defined
- Recording width: 50mm
- Printing speed: 25mm/s
- Recording type: Frozen waveform record
- NIBP recall record
- Trend table recordAlarm record
- Fixed- time record
ECG :
- Lead mode: 5 –leads (R,L,F,N,C)
- lead selection: I,II,III,avR,avL,avF,V
- Waveform: 2 and 6 channel selectable
- Gain selection: 0.25mm/mv, 0.5mm/mv, 1 mm/mv, 2 mm/mv
- Sweep speed: 12.5mm/s;25mm/s;50mm/s
- Heart Rate range: Adult: 15~300bpm;Neonate:/Pediatric: 15~350bpm
- Accuracy: ±1bpm or ±1%, whichever is greater
- Resolution: 1bpm
- Filter: Surgery mode :1~20Hz
- Monitor mode: 0.5~40Hz
- Diagnostic mode:0.05~130Hz
- Scaling signal: 1mV, ±3%
- Protection: Withstand 4000VAC/50 voltage isolation
- Against electrosurgical interference and defibrillation
- Alarm range: 15~350bpm
- S-T segment detection: Measurement range:-2.0mV~ +2.0mV
- Alarm Range: -2.0mV~ +2.0mV
- Accuracy: -0.8mV ~+0.8mV
- Error: ±0.02mV
- Arrhythmia analysis: Yes. 15types
NIBP :
- Method: Digital Automatic oscillometric
- Operation mode: Manual/Automatic/Continuous
- Auto measurement time: Adjustable(1~480min)
- Measurement Unit: mmHg/Kpa selectable
- Measurement types: Systolic, Diastolic, Mean
- Measurement range:
- Range of Systolic pressure: Adult: 40~270mmHg
- Pediatric:40~200mmHg
- Neonate: 40~135mmHg
- Range of mean pressure: Adult: 20~235mmHg
- Pediatric:20~165mmHg
- Neonate: 20~110mmHg
- Range of diastolic pressure: Adult: 10~215mmHg
- Pediatric:10~150mmHg
- Neonate: 10~100mmHg
- Over-pressure protection: Double safety
- Protection Resolution: 1mm Hg
- Alarm: Systolic, Diastolic, Mean
SPO2 :
- Measurement range: 0~100%
- Resolution: 1%
- Accuracy: ±2%(70-100%);0-69% unspecified
- Alarm range: 0~100%
- Pulse Rate: Range: 20~300bpm
- Solution: 1bpm
- Error: ±1bpm or ±2%, whichever is greater
RESPIRATION :
- Method: Thoracic impedance
- Measurement range: Adult : 7~120rpm; Neonate/Pediatric: 7~150rpm
- Apnea alarm: YES, 10~40s
- Solution : 1rpm
- Accuracy: ±2rpm
Temperature :
- Compatible probe: YSI or CYF
- Measurement Range: 0~50 C
- Resolution: 0.1 C
- Accuracy: ±0.1 C
- Channel: dual-channel, provide T1;T2,DT
- Refreshing time: about 1s
- Average measuring time: < 10s
IBP :
- Measurement range: -10~300mmHg
- Pressure name: ART,PA,CVP,RAP,LAP,ICP,P1,P2
- Channel: 2 channels
- Pressure transducer: sensitivity:5μV/VmmHg
- Impedance range: 300-3000Ω
- Accuracy: ±1mmHg or ±2%,whichever is greater(exclusive or transducer)
- Alarm Range: -10~300mmHg
EtCO2 :
- Measurement range: 0-99mmHg
- Accuracy: ±2mmHg(0~40mmHg)
- Sampling Range: 100ml/min
- Sampling rate accuracy: 15%
- Respiration rate: 0~120rpm
- Respiration accuracy: ±2rpm(0~70rmp) ±5rpm(>70rmp)
- Respiration time: < 240msec (10% to 90%)
- Delay time: < 2s
Standard Configuration :
- ECG
- Heart Rate
- Pulse Rate
- NIBP
- RESP
- Temp
- SPO2
- Printer
Optional : Thermal Recorder, 2-IBP, EtCO2 (side stream or main stream)
Safety
- IEC60601-1 approved, Category I BF, CF Type
Dimension and Weight
- Dimension 325mm x 400mmx 188mm
- Weight (with 2 batteries): < 7.0KGS
Operation Environment
- Temperature: Working 0~+50ºC
- Transportation and storage –20~+60ºC
- Humidity: Working ≤85%
- Transportation and storage ≤93%
- Power: AC100-240V,50/60Hz
|