Infant Equipment

MM-IC002 Infant Radiant Warmer

Product Description

Introduction:

Servo-controlled by micro-computer with option of automatic control or manual control the temperature; skin temperature sensor and bed temperature sensor; touch switch panel, setting temperature, real temperature, and APGAR timer can display separately, heating power rate indicates 0-100%; multi-function self-checking alarms; horizon angle of radiant head can be adjusted in two ways, there are 2pcs of radiant boxes; inclination of bassinet is adjustable; x-ray cassette is under the bed; 2pcs of oxygen cylinders and oxygen infusion device; RS232 connector, data output.

Standard Configure:

Radiation head, blue light radiant box (2pcs), temperature controller, skin temperature sensor, bed temperature sensor, infant bed, I.V. Pole, tray, x-ray cassette, oxygen cylinders, oxygen infusion device, low-pressure aspirator, resuscitation bag, infant head fixing frame, Stand.

Technical Specification:

  • Power supply : AC220V 50HZ
  • Power input: 950VA
  • Temperature control range25~37
  • Sensor precision: ≤0.3
  • Uniformity of bed temperature: ≤2.0
  • Adjustment angle for the radiation head 0°,30°,60
  • inclination of infant bed: ±15°
  • APGAR Timer: it has sound indication in 1min, 5min, 10min.

Alarming:

With audible and visible alarm, and it can indicate the real temperature and cut off the power supply of heater;
  • Over-temp alarm: when the skin temperature >39.5
  • Deviation alarm: indication temperature deviates away the setting temperature ±1;
  • Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the outside of bed.
  • Power failure alarm: when start up the unit, but there is no power and the power cord is disconnected.
  • System alarm: when there is system malfunction inside of temperature controller.

Package

  • Gross Weight150Kg
  • Net Weight113Kg
  • Measurement1270mm×810mm×1130mm