We offer durable, reliable and especially safe Infant Equipment because we understand the needs of small soul and cater them some special care through our infant care equipment. Our vision of infant equipment is considerable for extra care for infants which can be availed at very reasonable prices. Customers are also eased with the availability of the Infant Radiant Warmer to meet the varied requisites. In addition, we are the preeminent Infant Incubators Manufacturers Suppliers, reputed at Global level.
Features
- Tow control modes: Air mode and baby mode controlled by micro-computer
- The information can be stored, consulted and indicated by the LCD display screen
- Humidity and oxygen concentration display and servo control.
- The air circle with outlet in both front and back
- >37C temperature set function
- Double wall hood, air circulation system in the way of air outlet from front and back
- X-ray cassette under the bassinet
- Embedded integrated sensor module, integrated water reservoir in the shape of drawer
- Inclination of bassinet is adjustable
- Inclination of bassinet is adjustable
- The height of whole unit can be adjusted
- Failure alarm indication
- RS-232 connector
- Weighing system and Monitor shelf are optional
Standard configuration
- Baby compartment, main body, vertical height adjustment stand (VHA stand), the controller, I.V. Pole and shelf
- Weighing system and Monitor shelf are optional
Specification
- Power supply: AC220V-230V/50Hz
- Max power output: 1300VA
- Mode of control: Air mode and baby mode controlled by micro-computer
- Air temperature control range: 25°C-37°C, (override mode: 37-39°C)
- Baby temperature control range: 34-37°C, (override mode: 37-38°C)
- Accuracy of skin temperature sensor: ≤0.3°C
- Temperature variability: ≤0.5°C
- Temperature uniformity: ≤0.8°C (mattress is on the horizontal position)
- Inclination of the bassinet: ±8°
- Noise inside hood: ≤55dB (A) (Environment noise≤45Db(A))
- Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm, sensor box position alarm, tank position alarm, system alarm and so on
- Humidity display range: 0%RH~99%RH
- Humidity control range: 0%RH~90%RH
- Accuracy of humidity: ±10%RH
- Indication range of oxygen concentration: 0%~99%
- Control range of oxygen concentration: 20%~60%
- Accuracy of oxygen concentration: ±3%
- Range of weight: 100-8000g
- Resolution for weight indication: 1g
- Accuracy of weight: ±1%
Environmental temperature
- Operating range: 20°C~30°C (Set temperature should be higher 3°C than the environment temperature)
- Ambient air movement rate: ≤0.3m/s
Package
- Each unit is packed as two cartons
- Size (main body): 111cm x 71cm x 88.8cm, (cabinet): 103cm x 70cm x 83cm
- Gross weight (main body): 70Kg, (cabinet): 80Kg