Syringe Infusion Pump

MM-SIP004 Automatic Four Channels Syringe Pump

Product Description

Features:

  • Four Channels, sub-rate injection.
  • Four kinds of injection modes: Rate mode, Time mode, Body Weight mode, link mode.
  • Innovative design of outer appearance gives a strong sense of science and technology.
  • Double CPU makes the process of injection safer and more reliable.
  • Brand-new STM32 ARM-Cortex M3 plat-form, optimized circuit design, operation rate more faster and stable.
  • The use of TFT color LCD makes the screen display more remarkable.
  • Digital menu, easy operation; functions fully meet the clinical demand.
  • Ample and user-friendly indicator information; accurate and complete alarm information to ensure safe injection.
  • Aluminum alloy base, integrative pump device makes more smooth operation and easier installation and maintenance.


Specifications


Maximum injection rate : 1500ml/h (Different injectors have different maximum rate.)
Range of flow rate : 5ml injector: 0.1-100ml/h
10ml injector: 0.1-200ml/h
20ml injector: 0.1-400ml/h
30ml injector: 0.1-600ml/h
50ml injector: 0.1-1500ml/h
Injection increment : 0.1ml/h
Bolus rate : 5ml injector: 100ml/h
10ml injector: 200ml/h
20ml injector: 400ml/h
30ml injector: 400-600ml/h
50ml injector: 400-1500ml/h
Injection mode selection : 1. Rate mode 2. Time mode 3. Body Weight mode
Rate mode : Speed: 0.1-1500ml/h (maximum value depends on the injector specifications)
Time mode : Time:1-2000 minutes
Accumulated injection volume: 0.1-999.9ml
Body Weight mode : Weight: 0.1-300.0kg
Drug: 0.1-999.9(mg)
Volume: 0.1-999.9ml
Dose: 0.1-9999.9(restricted to unit, drug, solution, weight and injector specification).
Unit: mg/kg/h, ug/kg/min
KVO rate : Adjustable range: 0.1ml/h-5.0ml/h (begin the KVO function under the occlusion situation) ; pressing STOP key to cancel the KVO function when it is unnecessary in use.
IP grade : IP21
Injection pressure : The maximum pressure is 0.3Mpa, Occlusion alarm pressure value is 40kPa-160kPa when pressure value is minimum, injection rate is 25ml/h, the longest alarm time is 10 minutes. When pressure value is minimum, injection rate is 5ml/h, the longest alarm time is 10 minutes.
Preset volume : 0.1ml-9999.9ml
Preset time : 0-99 hours, 0-59 minutes
Accumulated injection volume : 0.1ml-9999.9ml
Accuracy : 3%
Power supply : AC 100-220V,50/60Hz
Battery : Rechargeable lithium polymer battery, 7.4 V, 1600mAh.
Maximum power consumption : 25VA, running more than 2 hours at the rate of 25ml/h after being fully charged.
Battery charge : When syringe pump is connected to AC power, the battery will automatically recharge. (about 8-14 hours to recharge fully.)
Fuses : T2AL250V-
Displayed information : Flow rate; accumulated injection volume; injector specification; battery capacity; bed No.; AC power indicator, operation indication, etc.
State indication information : Stop, inject, bolus, KVO (stop flashing, the other indicators flash by turns.)
Alarm information : near completion, Over, occlusion, low battery, abnormal 1 (communications failure), abnormal 2 (pump stuck), abnormal 3 (wrong parameter), volume limited alarm, no AC power
Maximum size of the outer shell : 288x130x122 mm ( lengthx width x height )
Maximum weight : <3.0kg
Classification : ClassII, Type BF
Shell material : ABS plastic
Operating conditions : Environment temperature 5-40, air pressure 80kPa -106kPa,relative humidity 80%
Storage conditions : Environment temperature -40-55, air pressure 50kPa -106kPa,relative humidity 95%
Applicable syringes : 5ml.10 ml.20 ml.30ml.50 ml syringes
Electrical safety : In accordance with the stipulation of GB9706.1-1995
Applied standards : According to the stipulation of YZB/Yue 001-2008